| Wavelength
|
 |
 |
| For
the measurement of this fundamental parameter, Starna® offers you
the choice of either sealed cell, or filter materials. Please note
that your selection process should not only include the wavelength(s)
required, but also the spectral bandwidth (SBW) of your spectrophotometer.
Use of the reference materials outside the SBW range detailed is not
excluded, but at higher spectral bandwidths some of the certified
peaks may not be resolved from the overall background spectrum. |
Holmium
Oxide - UV and Visible Wavelength
Description: Holmium
oxide (4% m/v) in 10% v/v perchloric acid.
Primary
Usage: Assessment
of wavelength scale accuracy in both UV and visible regions.
Useable
range:
240 nm to 650 nm, instruments with spectral bandwidth of less than
3 nm.
Physical
Configuration: Far UV quartz
cells that have been permanently heat sealed. |
Holmium
Oxide - UV and Visible Wavelength
The use of holmium oxide solvated in perchloric acid
is an established and well recognised method for the validation of
the wavelength scale of a spectrophotometer in the UV and visible
regions. When prepared in perchloric acid, holmium oxide gives a spectral
scan containing a series of characteristic peaks. However, these values
are dependent and will vary with the spectral bandwidth of the measuring
instrument (NIST Special Publication 260-120)
RM-HL: (see
MSDS)
Consists of one sealed cell, with certified peak at
spectral bandwidth values of 0.10 nm, 0.25 nm, 0.50 nm, 1.00 nm, 1.50
nm, 2.00 nm and 3.00 nm. |
Typical
values obtained:
| SBW
|
nm
|
nm
|
nm
|
nm
|
nm
|
nm
|
nm
|
nm
|
nm
|
nm
|
nm
|
nm
|
nm
|
nm
|
| 0.10
|
640.54 |
536.47 |
485.26 |
467.88 |
452.13 |
416.16 |
385.55 |
361.43 |
345.60 |
333.52 |
287.10 |
278.29 |
249.85 |
241.07 |
| 0.25
|
640.52 |
536.52 |
485.29 |
467.84 |
452.12 |
416.16
|
385.48 |
361.42 |
345.46 |
333.56 |
287.12 |
278.26 |
249.88 |
241.09 |
| 0.50 |
640.52
|
536.55 |
485.31 |
467.90
|
451.99 |
416.19 |
385.61 |
361.39 |
345.54 |
333.61 |
287.19 |
278.29 |
249.94 |
241.6 |
| 1.00
|
640.65
|
536.85 |
485.37 |
467.93 |
451.38 |
416.58 |
385.90 |
361.18 |
345.56 |
333.64 |
287.66 |
278.28 |
250.14 |
241.33 |
| 1.50 |
640.58
|
536.66 |
485.31 |
467.88 |
451.37 |
416.40 |
385.78 |
361.41 |
345.53 |
333.58 |
287.39 |
278.27 |
250.00 |
241.30 |
| 2.00 |
640.85
|
537.01 |
485.29 |
468.00 |
451.42 |
416.73 |
385.90 |
361.13 |
345.66 |
333.62 |
287.78 |
278.25 |
250.25 |
241.34 |
| 3.00 |
641.29 |
537.50 |
485.28 |
468.14
|
451.45 |
417.04
|
386.17 |
361.11 |
345.67 |
333.74 |
287.91 |
278.30 |
250.38 |
241.34 |

|
 |
• Additional traceability
links to NIST primary materials are established using SRM 2034 holmium
oxide (4% m/v) in perchloric acid (10% v/v).
•Use: All appropriate fundamental parameters
and procedures relating to measurement, handling and storage are
fully documented on the certificate supplied with each Certified
Reference Material.
|
|