Didymium
Glass -UV and Visible Wavelength
Description: Didymium
glass filter.
Primary
Usage: Assessment of wavelength scale accuracy
in the visible/near NIR region.
Useable
range: 430 nm to 890 nm, instruments with
SBW of less than 10 nm.
Physical
Configuration: Glass filters 'stress free' mounted in anodised
aluminium holder. |
 |
Product
Description:
Like its liquid counterpart, the didymium glass
filter produces characteristic peaks that make it suitable for use
as a wavelength reference material in the high end visible/near NIR
region. As a solid material it is physically more robust than the
liquid cell, and can therefore be used in a more demanding environment,
providing care is still taken to avoid scratching the surface of the
filter. By individually certifying each Starna®
filter, uncertainty caused by variations in peak positions (observed
from melt to melt of the glass)
is significantly reduced to acceptable levels (± 0.2 nm). (see
MSDS)
RM-DG:
Consists of one filter, "stress free" mount in a proprietary
NIST design, with certified peak at spectral
bandwidth values of 0.10 nm, 0.25 nm, 0.50 nm, 1.00 nm, 1.50 nm, 2.00
nm, 3.00nm. |